Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy and error. The probe
compensation procedure is as follows:
1. Perform step 1, 2, 3 and 4 of “Function Inspection” in the previous section.
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-10 Probe Compensation
3. Use a nonmetallic driver to adjust the variable capacitor on the probe until the
displayed waveform is as the Correctly compensated in the figure above.
Over compensated Correctly compensated Under compensated
Vue de la page 33