RIGOL Chapter 2 Performance Test
Service Guide for DM3068
2. 5 Routine Test
2.5.1 Zero-offset Test
The zero-offset test inspects the meter’s offset performance at the point of zero. It is
necessary only when a regular offset is occurred to a function or range.
To perform a zero-offset test:
1. Make sure you have carefully read the “Test Conditions”.
2. Use a 4-terminal short-circuiter to short connect the HI-LO and Sense HI-LO for
DC voltage and resistance measurements (see figure below) and open both
200mA and 10A current input terminals in current measurement.
3. Perform a step-by-step test for the functions under different ranges specified in
the following table and set the integration time to 100PLC. Please note that none
of the math function is allowable unless otherwise stated.
4. Compare the tested results with the limits in the table.
Figure 2-2 Zero-offset test connection
Table2-3 Zero-offset test items