Rigol DS1000D Manuel d'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 137
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
Command Systems RIGOL
Programming Guide for DS1000E, DS1000D Series
2-25
Example:
:TRIG:HOLD 0.0005 Setup the holdoff time as 500ms.
:TRIG:HOLDoff? The query returns 5.000e-04
7. :TRIGger:STATus?
Command Format:
:TRIGger:STATus?
Function Explanation:
The command queries the operating status of the oscilloscope. The status could
be RUN, STOP, T`D, WAIT or AUTO.
Returned Format:
The query returns RUN, STOP, T'D, WAIT or AUTO.
8. :Trig%50
Command Format:
:Trig%50
Function Explanation:
The command sets the trigger level to the vertical midpoint of amplitude.
9. :FORCetrig
Command Format:
:FORCetrig
Function Explanation:
The command forces the oscilloscope to trigger signal, which is usually used in
“Normal” and “Single” mode.
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 136 137

Commentaires sur ces manuels

Pas de commentaire