RIGOL Chapter 15 System Function Setting
MSO2000A/DS2000A Users Guide
15-16
Aux Output
Users can set the type of the signal output from the [Trigger Out] connector at the
rear panel.
Press Utility AuxOutput to select the desired output type.
TrigOut
After this type is selected, the oscilloscope outputs a signal that can reflect the
current capture rate of the oscilloscope at each trigger. Connect this signal to a
waveform display device, measure the signal frequency and the measurement
result equals the current acquisition rate.
PassFail
After this type is selected, the oscilloscope will output a pulse signal from the
[Trigger Out] connector at the rear panel when failed waveforms are detected.