Figure 7 To Use the Compensation Signal
4. Press the AUTO key.
5. Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
Figure 8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform “Probe
Compensation” in the next section.
WARNING
T
o avoid electric shock during the use of probe, please make sure that
the insulated wire of the probe is in good condition and do not touch
the metallic part of the probe when the probe is connected to high
voltage source.
The signal output from the probe compensation connector can only be used for
probe compensation adjustment and can not be used for calibration.
Compensation Signal Output Terminal
Ground Terminal