RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
2-54
In the trigger System, there is a general command to the edge, pulse width, video,
slope and alternate trigger.
1. :TRIGger<mode_1>[<mode_2>]:SENSitivity
Command Format:
:TRIGger<mode_1>[<mode_2>]:SENSitivity<count>
Explanation:
The command is to set the trigger sensitivity from 0.1~1div.
<mode_1> can be :EDGE, :PULSE, :VIDEO, :SLOPE and :ALTERNATION;
When <mode_1> is :ALTERNATION, [<mode_2>] can be :EDGE, :PULSE,
:VIDEO or :SLOPE, else omit [<mode_2>].
Query Format:
:TRIGger<mode_1>[<mode_2>]:SENSitivity?
Returned Format:
Query returns the set value of <Count> in scientific numeric notation.
Example:
:TRIGger:EDGE:SENSitivity0.2 Set the edge trigger sensitivity as 0.2div
:TRIGger:EDGE:SENSitivity? Query returns 2.00e-01