Users Guide for DS1000B series
2-102
Track Mode
Press CursorModeTrack to display the manual mode menu.
Figure 2-142 Table 2-87 Cursor descriptions
Menu
Settings
Comments
Mode
Track
Set Track mode in cursor measurement.
Cursor A
CH1
CH2
CH3
CH4
MATH
None
Set Cursor A in conjunction with CH1.
Set Cursor A in conjunction with CH2.
Set Cursor A in conjunction with CH3.
Set Cursor A in conjunction with CH4.
Set Cursor A in conjunction with MATH.
Turn off Cursor A i
Cursor B
CH1
CH2
CH3
CH4
MATH
None
Set Cursor B in conjunction with CH1.
Set Cursor B in conjunction with CH2.
Set Cursor B in conjunction with CH3.
Set Cursor B in conjunction with CH4.
Set Cursor B in conjunction with MATH.
Turn off Cursor B.
CurA
(Cursor A)
Turn the multifunctional Knob ( ) to
move cursor A Horizontally.
CurB
(Cursor B)
Turn the multifunctional knob ( ) to
move cursor B horizontally.
In cursor track mode, the cursors move together with the selected waveform.
To do Track mode Cursor Measurements, follow these steps:
1. Select the Track mode for cursor measurement by pressing soft button as:
CursorModeTrack.
2. Select channel Source for Cursor A and Cursor B by pressing soft button as:
CursorCursor A or Cursor BCH1, CH2, CH3, CH4, MATH or None.
3. Move the cursors to adjust the horizontal positions of the cursors: (Details in the
following Table )