RIGOL Command Systems
Programming Guide for DS1000B Series
2-62
13. :MEASure:RISetime?
Command Format:
:MEASure:RISetime? [<source>]
Function
The command measures the rise time of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 4.000e-005, and the unit is s.
14. :MEASure:FALLtime?
Command Format:
:MEASure:FALLtime? [<source>]
Function:
The command measures the fall time of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 4.000e-005, and the unit is s.
15. :MEASure:PERiod?
Command Format:
:MEASure:PERiod? [<source>]
Function:
The command measures the period of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 1.000e-003, and the unit is s.