RIGOL Chapter 2 Front Panel Operations
User’s Guide for DM3068
2-40
Measurement Configuration
Most measurement parameters are user-defined. Changing a measurement
parameter will change the measurement precision and speed as well as the input
impedance. An appropriate measurement parameter based on the actual
application will ensure faster measurement or higher measurement precision.
The default measurement configurations of the multimeter can ensure the
accuracy of the measurement results in most cases. Users can directly use these
defaults for any measurement or modify the parameters of the measurement
function as required.
The parameters for different measurement function differ, see table below.
Table 2-5 Measurement parameter
Functions Parameters
DCV integration time, DC impedance, Auto zero (AZ)
ACV AC filter
DCI integration time, Auto zero (AZ)
ACI AC filter
OHM (2WR, 4WR) Integration time, Auto zero (AZ),
Offset compensation
(OC)
CAP N/A
CONT Short-circuit resistance
DIODE Current Source
FREQ/PREIOD Gate time, AC filter
SENSOR N/A
To set the measurement parameter, select a measurement function and then press
. For more details, please refer to the following content.