9
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy and error. The probe
compensation procedures are as follows:
1. Perform step 1, 2, 3 and 4 of Function Inspection.
2. Check the displayed waveforms and compare them with the following figures.
Figure 8 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is as the “Correctly
compensatedin the figure above.
Over compensated Perfectly compensated Under compensated