User’s Guide RIGOL Publication number: UGC01107-1110 Dec. 2008 DM3000 Series Digital Multimeter DM3061/2
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series VIII Content Safety Notices ...
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-50 Press Log Stop, the interface shows below. Figure 2- 59 T
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-52 2. The Data Log Trigger and Delay To set the start condition and
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-54 3. The Data log stop To set the conditions to stop Data Log. P
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-56 Multi-Route Scanning DM3054 and DM3064 have a built-in inspection
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-58 Table 2- 41 The Scanning Menu Description Function Menu Setting D
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-60 Figure 2- 74 Table 2- 43 The Task Menu Description Fun
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-62 Edit the Scan Task Press Edit in figure 2-68, enter the task edi
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-64 How to Use the Built-in Help System The built-in help system prov
RIGOL
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-4 4. Start Statistical measurement. (1). Press ON, turn on the Stat
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-6 Example 3: dBm Measurement Do these steps as follow: 1. Connect te
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-8 8. Lead test leads into circuit and start to measure. Figure 3- 8
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-10 Example 6: Temperature Sensor Setting a sensor is the same way of
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-12 c) Press Unit ℃, select the unit of the sensor: ℃. Figure 3- 2
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 3-14 b) Current sensor: Figure 3- 26 Sensor
RIGOL
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 4-2 12. DCV is -1200V to 1200V In limit test and null value settings th
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 4-4 To Change the Electric Power Fuse The electric power fuse located i
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-2 [1] Specifications are for 60 minutes warm–up and select 6½ reading r
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-4 [1] Specifications are for 60 minute warm–up and select 6½ reading res
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-2 General Inspection Inspect a new DM3000 Digital Multimeter with th
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-6 Capacitance Characteristics Accuracy Specifications (% of reading + % of
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-8 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-10 Display: 256 x 64 pixels LCD
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-12 [1] Specifications are for 60 minute warm–up, selecting 5 3/4 reading
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-14 [1] Specifications are for 60 minute warm–up and select 5 3/4 reading
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-16 Capacitance Characteristics Accuracy Specificat
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-18 Shunt Resistor: 0.025Ω for 1A, 10A, 1.025Ω for 200mA, 11.025Ω for
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 5-20 General Specifications Display:
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 6-2 Appendix B: Warranty RIGOL warrants that the products that it manufact
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 6-4 Appendix D: Contact RIGOL If you have any problem or requirement oc
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-4 The Front/Rear Panel and User Interface It is important to get fam
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-6 To Measure DC Voltage The following shows the system connections a
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-8 To Measure AC Voltage The following shows the system connections a
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-10 To Measure DC Current The following shows the system connections
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-12 To Measure AC Current The following shows the system connections
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-14 To Measure Resistance The following shows the system connections
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-16 4-Wire Resistance Measurement Figure 1- 21 Table1- 6 R
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-18 To Measure Capacitance The following shows the system connections
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-20 To Test Continuity The following shows the system connection
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-22 To Check Diodes The following shows the system connections and se
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-24 To Measure Frequency and Period The following shows the system co
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-26 Period Test Figure 1- 34 Period Measurement Data Interface
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series II Safety Notices Review the following safety precau
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-28 To Measure Sensor The DM3000 converts the sensor physical p
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-30 Press Add button, to input the Measured and Corresponding value to
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-32 Figure 1- 53 The History Data Use the history function to review
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-34 To Choose Reading Resolution The measurement reading resolutions
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-36 To Choose Range Options Use “manual” selection or “automatic”
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 1-38 To Control Trigger Options Use or to trigger the multimeter
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-2 To Set up Measurement Parameters Press key to enter the Mea
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-4 AC Filter There are three settings at the AC Filter menu. Choose th
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-6 Null Measurement The DM3000 null settings is available for DC volta
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-8 Ratio Measurement Use Ratio measurement to measure the ratio
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series IV Safety Terms and Symbols Terms in This Guide. Thes
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-10 Frequency Measurement Frequency measurement function is used for m
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-12 Math Functions Press key, the display shows: Figure 2- 11 The
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-14 Math Functions Selective The DM3000 provides five math functions:
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-16 2. Limit Measurement Use the Limit test function to perform pass/
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-18 4. dBm Measurement This function applies to AC voltage and DC vol
RIGOL
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-20 Auto Triggering Auto triggering takes continuous measurement at th
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-22 Reading Hold Function Start the Reading Hold Function, the hold me
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-24 External Triggering is used to set the parameter which initiates
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-26 To Set up the VMC At external triggering mode, when the data sampl
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-28 Local/U-Disk Storage Local storage block is built-in the multimete
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series VI RIGOL DM3000-Series Digital Multimeter is a
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-30 Figure 2- 30 NOTE Store, recall and delete use the same interf
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-32 To Set Up the Utility The Utility function establishes system
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-34 Set Up LAN I/O Parameter LAN Parameters Following the procedures
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-36 DNS Settings Press I/O LAN DNS, the display shows: Figur
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-38 Set Up RS-232 I/O Parameters Choose RS-232 (serial data interface)
RIGOL
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-40 Select languages DM3000 supports two kinds of languages for users.
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-42 Beeper On/Off Press Sys Sound, enter the menu shown below:
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-44 Set Up Digit Format Press Sys Format, and enter the m
RIGOL
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-46 Factory Defaults Setting Press System Cfg Default, then r
RIGOL © 2007 RIGOL Technologies, Inc. User‟s Guide for DM3000 Series 2-48 High-Speed Data Log High-speed data log features: display mode s