RIGOL Chapter 1 Quick Start
User’s Guide for DSA800 Series
1-12
3. GEN OUTPUT 50Ω
The output of the tracking generator can be connected to a receiver through a
cable with an N male connector. The tracking generator is only available for
DSA815-TG/DSA832-TG/DSA875-TG.
CAUTION
To avoid damage to the tracking generator, the reverse power cannot
exceed +10 dBm when the frequency is lower than 10 MHz; the reverse
power cannot exceed +20 dBm when the frequency is greater than 10
MHz. The reverse DC voltage cannot exceed 50 V.
4. RF INPUT 5
The input terminal of the signal under measurement. [RF INPUT 50Ω] can
be connected to the device under measurement via a cable with an N male
connector.
CAUTION
To avoid damage to the instrument, for the signal input from the RF
input terminal, the DC voltage component and the maximum continuous
power of the AC (RF) signal component can not exceed 50 V and +20
dBm respectively.