Rigol DSA875 Manuel d'utilisateur Page 80

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 235
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 79
RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA800 Series
2-28
Sweep/Trig
Set parameters about the Sweep and Trigger functions, including sweep time, auto
sweep time, sweep mode, number of sweep, trigger type, etc.
Sweep Time
Set the time needed for the spectrum analyzer to finish a sweep within the span
range. The sweep time can be set in “Auto” or “Manual” mode and the default is
Auto”.
Key Points:
In non-zero span, the analyzer selects the shortest sweep time on the basis of
the current RBW and VBW settings if Auto is selected.
Decreasing the sweep time would speed the measurement. However, an error
may be caused if the specified sweep time is less than the minimum sweep
time in Auto coupling; at this point, UNCALis shown in the status bar on the
screen.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer toParameter Setting.
Table 2-16 Sweep Time
Parameter
Explanation
Default
37.5 ms
Range* 20 us to 7500 s
Unit ks, s, ms, us, ns, ps
Knob Step Sweep time/100, Min = 1 us
Direction Key Step in 1, 1.5, 2, 3, 5, 7.5 sequence
Note: *The minimum in non-zero span mode is 1 ms.
Vue de la page 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 234 235

Commentaires sur ces manuels

Pas de commentaire