Chapter 2 Basic Waveform Output RIGOL
User’s Guide for DG5000
2-7
To Output Pulse Waveform
To output a pulse waveform from CH1 with 20 kHz frequency, 2.5 Vpp amplitude, 500
mV
DC offset, 10 μs pulse width (20% duty cycle), 50 ns leading and trailing edge time
and 8 μs delay. Refer to To Output Sine Waveformto configure the parameters
and output. This section will only talk about Width/Duty, Leading, Trailing,
Delayand Restore.
Pulse Width/Duty Cycle
Pulse Width is defined as the time from the 50% threshold of a leading edge
amplitude to the 50% threshold of the next trailing edge amplitude as shown in the
figure above. Duty Cycle is defined as the percentage that the pulse width takes up in
the pulse period.
Press Pulse Width/Duty to highlightPulse Width, and then use the numeric
keyboard to input 10and select the unitμsecin the pop-up menu.
Pulse Width is limited by the Minimum Pulse Width” and the “Pulse Period”.
Minimum Pulse Width = 4 ns
Pulse Width Minimum Pulse Width
Pulse Width Pulse Period - Minimum Pulse Width × 2
Pulse Width and Duty Cycle are correlative. Once a parameter is changed, the
Pulse Width
10%
Leading Edge Time
Trailing Edge Time
90%
50%
Pulse Period