RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
Accuracy, Average
Acquisition Mode
±(DC Gain Accuracy× reading+0.1div+1mV)
Average of ≥16 Waveforms with vertical position not at zero:
±[DC Gain Accuracy× (reading+vertical position)+(1% of
vertical position) + 0.2div]
Add 2mV for settings from 2mV/div to 245 mV/div
Add 50mV for settings from 250mV/div to 10V/div
Delta Volts
Measurement
Accuracy (Average
Acquisition Mode)
Delta Volts between any two averages of 16 waveforms
acquired under same setup and ambient
conditions: ± (DC Gain Accuracy× reading + 0.05 div)
0.1div~1.0div (adjustable)
± 6 divisions from center of screen
Trigger Level Accuracy
(typical) applicable for
the signal of rising
and falling time ≥20ns
± (0.3div × V/div)(± 4 divisions from center of
screen)
± (6% of setting + 200 mV)
Normal mode: pre-trigger (storage depth/ 2*sampling
rate), delayed trigger 1s
Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Set Level to 50%
(Typical)
Input signal frequency ≥50Hz
Rising, Falling, Rising + Falling
(>, <, =) positive pulse width, (>, <, =) negative pulse
width
Video standard &
line frequency
Support standard NTSC, PAL and SECAM broadcast systems.
Line number range: 1~525 (NTSC) and 1~625 (PAL/SECAM)
(>, <, =) Positive slope, (>, <, =) negative slope