User’s Guide for DS1000E, DS1000D Series
2-99
Manual Mode
Figure 2- 121 Table 2- 79 The Manual Mode menu
Menu
Settings
Comments
Mode
Manual
Adjust the cursor to measure X/Y
parameters manually
Type
X
Y
Shown as vertical line to measure the
horizontal parameters
Shown as horizontal line to measure
the vertical parameters
Source
CH1
CH2
MATH/
FFT
LA
Select the measurement signal source.
(LA is only for DS1000D series)
In this mode, the oscilloscope measures the Y or X coordinate values of the cursors,
and the increments between the two cursors.
To do manual Cursor Measurements, please do as the following steps:
1. Select the Manual mode for cursor measurement by pressing soft button as:
CursorModeManual.
2. Select channel Source for measurements by pressing soft button as:
CursorSourceCH1, CH2, MATH (FFT) or LA (DS1000D series)
NOTE: While measuring the channel of MATH, the results are valued with “d”
(division) as units.
3. Select the cursors type by pressing soft button as CursorTypeX or Y.
4. Move the cursors to adjust the increment between the cursors: (Details in the
following Table)