User’s Guide for DM3000 Series
3-15
Example 7: Reading Hold
Do these steps as follows:
1. To measure a DC Voltage. Connect test leads as shown below:
Figure 3- 27
2. Press
button, select DC Voltage measurement function. Choose an
appropriate measurement range.
3. Set the Hold measurement function parameters.
(1). Press Æ Auto Æ Hold Æ 0.1%, set up the hold scope to 0.1%.
(2). Press
, to save all changes, back to a higher level menu.
(3). Press
, to back to a higher level menu.
4. When the
was lit, press this button once, the button will blink, it means the
trigger mode is now Hold mode. If current trigger mode is single, press
button twice.
5. Lead test leads into circuit, start to measure.
Figure 3- 28
DC Voltage