RIGOL
User’s Guide for DM3058/DM3058E
To Check Diode
Operating Steps:
1. Press
to enter into the following interface.
Figure 2-28 Testing interface of Diode
2. Connect red lead to both terminal Input-HI and anode of the Diode and black
lead to both terminal Input-LO and cathode of the Diode as the figure below.
Figure 2-29 Sketch map for connecting
3. Check the connection state of Diode.
When connecting, the instrument will beep one time (If Sound is open).
Table 2-9 Check Diodes Characteristics
Test Current 1mA
Range Fixed at 2.0V
Open-circuit Voltage <8V
Input Protection 1000V (HI terminal)
Beep Condition 0.1V≤V
measured
≤2.0V