Rigol DM3058 Manuel d'utilisateur Page 48

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 138
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 47
RIGOL
Users Guide for DM3058/DM3058E
2-26
Table 2-10 Frequency Test Characteristics
Ranges 200mV, 2V, 20V, 200V, 750V
Input Signal Range 20Hz ~ 1MHz
Input Protection 750VRMS on all ranges (HI terminal)
Configurable Parameters REL
3. Set the relative value (Optional).
Press REL to open or close Relative operation function. When it is open, “REL
is shown over the screen and the reading displayed is a relative value which
comes from the result of Actual measurement value subtracts the value that
has been set. (For the details about how to set, please refer to “
Math
Functions” in Chapter 2.)
4. Read measurement value
Frequency measurement is fixed at “Slow” rate, 5.5 digits resolution. Therefore,
you cannot adjust the reading rate when reading the result.
5. View history data
Press History to enter into the following interface and check the data getting
from this measurement and saved.
Figure 2-32 View interface of History information
There are three types of way for viewing historical date: “Info”,List” and
HistoG”. Press Save after checking. By pressing Update, the History
information will be refreshed to be current newest information.
Vue de la page 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 137 138

Commentaires sur ces manuels

Pas de commentaire