Manual Mode
Press Cursor→Mode→Manual to display the manual mode menu.
Figure 2-115 Table 2-75 Manual mode menu
Adjust the cursor to measure X/Y
parameters manually.
Shown as vertical line to measure the
horizontal parameters.
Shown as horizontal line to measure the
vertical parameters.
Select the measurement signal source.
Set CurA effective and adjust it.
Set CurB effective and adjust it.
In this mode, the oscilloscope measures the Y or X coordinate values of the cursors,
and the increments between the two cursors.
To do manual Cursor Measurements, please do as the following steps:
1. Select the Manual mode for cursor measurement by pressing soft button as:
Cursor→Mode→Manual.
2. Select the cursors type by pressing soft button as Cursor→Type→ X or Y.
3. Select channel Source for measurements by pressing soft button as:
Cursor→Source→CH1, CH2 or MATH/FFT.
4. Move the cursors to adjust the increment between the cursors:(Details in the
following Table)