RIGOL
User‟s Guide for DS1000CA Series
2-65
3. Pass/Fail connection
The Pass/Fail connection has an optically isolated output. It needs to connect to
another circuit to fulfill the function.
Before connecting to an external circuit, make sure the maximum
voltage/current does not exceed the internal photo MOS relay‟s rating,
400V/100mA. DS1000CA series adopts the technique of Optical Isolation, and
the output device has no polarity and can be used in any circuit within the
ratings.
Figure 2-90
Schematic Diagram of Pass/Fail output