Programming Guide for DM3058/DM3058E
4-18
1. DATA:FEED RDG_STORE, {"CALCulate"|""}
Command Format:
DATA:FEED RDG_STORE, {"CALCulate"|""}
Function:
The command selects whether readings taken using the INITiate command are
stored in the multimeter’s internal memory (default) or not stored at all. Note that
the command does not execute any actual operation as restricted by working
principle.
2. DATA:FEED?
Command Format:
DATA:FEED?
Function:
The command queries the storage state of readings, the returned value is "CALC"
or "". Note that the query only returns “CALC” as restricted by working principle.
3. DATA:POINts?
Command Format:
DATA:POINts?
Function:
The command queries the number of readings in the internal memory of the
multimeter.