RIGOL
Programming Guide for DM3058/DM3058E
3-5
The command is valid only when AVERAGE operation is enabled.
Return Format:
The query returns the calculated average.
5. :CALCulate:STATistic:COUNt?
Command Format:
:CALCulate:STATistic:COUNt?
Function:
The command queries the number of current measurement calculated.
Explanations:
The command is only valid for current measurement. If measurement function is
changed, it will restart the counting.
Return Format:
The command queries the calculated number of measurement.
6. :CALCulate:STATistic:STATe
Command Format:
:CALCulate:STATistic:STATe?
:CALCulate:STATistic:STATe {ON|OFF|1|0}
Function:
The commands query and set the statistic state.
Explanations:
:CALCulate:STATistic:STATe?Queries the current statistic state.
:CALCulate:STATistic:STATe {ON|OFF|1|0} – Sets the desired statistic state.
Return Format:
The query returns the current statistic state ON (1) or OFF (0).
Example: