Function:
The commands set and query the precision of DC voltage measurement.
Parameter:
Name Type Range Default
<resolution> Discrete Refer to Table 4-2 1ppm x range
Explanations:
[SENSe:]VOLTage:DC:RESolution {<resolution>|MINimum|MAXimum} Sets
the desired precision of DC voltage measurement.
[SENSe:]VOLTage:DC:RESolution? [MINimum|MAXimum] Queries the current
precision of DC voltage measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.
Return Format:
The query returns the current precision of DC voltage measurement.
19. [SENSe:]VOLTage:AC:RESolution
Command Format:
[SENSe:]VOLTage:AC:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]VOLTage:AC:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of AC voltage measurement.
Parameter:
Name Type Range Default
<resolution> Discrete Refer to Table 4-2 1ppm x range
Explanations:
[SENSe:]VOLTage:AC:RESolution {<resolution>|MINimum|MAXimum} Sets the
desired precision of AC voltage measurement.
[SENSe:]VOLTage:AC:RESolution? [MINimum|MAXimum] Queries the current
precision of AC voltage measurement.
<resolution> relates to measurement rate and integration time, refer to Table