Chapter 6 MATH and Measurement RIGOL
MSO1000Z/DS1000Z User’s Guide 6-9
optimal value according to the current configuration.
Tips
You can adjust the center frequency and horizontal scale at the same time
using HORIZONTAL
SCALE.
Signals with DC components or deviation would cause error or deviation of the
FFT waveform components. To reduce the DC components, set the "Channel
Coupling" to "AC".
To reduce the random noise and aliasing frequency components of repetitive
or single pulse, set the "Acquisition Mode" of the oscilloscope to "Average".