RIGOL Chapter 15 Accessibility Setting
15-8 MSO1000Z/DS1000Z User’s Guide
Power-off Recall
You can set the system configuration to be recalled when the oscilloscope is powered
on again after power-off. Press Utility System Power Set to select "Last" or
"Default".
Last: return to the setting of the system at last power-off.
Default: return to the factory setting of the system.
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channel connections, and then press Utility Self-Cal and
the self-calibration interface as shown in the figure below is displayed.
Figure 15-2 Self-calibration Interface
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note: Most of the keys are disabled during the self-calibration.