Chapter 6 MATH and Measurement RIGOL
MSO1000Z/DS1000Z User’s Guide 6-25
Delay and Phase
Source 1
Source
2
Period
Delay
Figure 6-4 Delay and Phase
Source 1 and source 2, the same as source A and source B in the measurement
setting menu can be any channel of CH1-CH4 or D0-D15. For the setting method,
please refer to the introduction in "Measurement Setting
".
1. Delay 12 : the time difference between the rising edges of source 1 and
source 2. Negative delay indicates that the selected rising edge of source 1
occurred after that of source 2.
2. Delay 12 : the time difference between the falling edges of source 1 and
source 2. Negative delay indicates that the selected falling edge of source 1
occurred after that of source 2.
3. Phase 12 : phase difference calculated according to "Delay 12 " and the
period of source 1, expressed in degree. The calculation formula is as shown in
equation (6-1).
4. Phase 12 : phase difference calculated according to "Delay 12 " and the
period of source 1, expressed in degree. The calculation formula is as shown in
equation (6-1).
The phase calculation formula:
×360
Delay
Phase =
Period1
°
(6-1)
Wherein,
Phase
represents "Phase 12 " or "Phase 12 "
Delay
represents "Delay 12 " or "Delay 12 "
1Period
represents the period of source 1