Rigol MSO/DS1000Z Series Manuel d'utilisateur Page 176

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 270
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 175
RIGOL Chapter 8 Protocol Decoding
8-14 MSO1000Z/DS1000Z User’s Guide
Note: By default, MSO1000Z/DS1000Z uses "Trace" as the data source.
9. Address Information during Decoding
In I2C bus, the front part of each frame of data contains the address information
and blue patches are used to represent the address ID. In the ID, "Write" is used
to represent writing address and "Read" is used to represent reading address.
Press ADDR to select "Normal" or "R/W". When "R/W" is selected, "AddrBits"
will include the "R/W" bit as a part of the address value.
Figure 8-6 I2C Address Information during Decoding
When the ACK (ACKnowledge Character) is not met, error marks "?" as shown in
the figure below will be displayed.
Figure 8-7 I2C Error Expressions during Decoding
ACK=1
Address ID
Vue de la page 175
1 2 ... 171 172 173 174 175 176 177 178 179 180 181 ... 269 270

Commentaires sur ces manuels

Pas de commentaire