RIGOL
2-74 © Copyright RIGOL Technologies, Inc. 2007.
User’s Guide for DS1000 Series
Figure 2-88 Table 2-54
Menu Settings Comments
Pass/Fail Setup Pass/Fail test
Record Setup Waveform Recorder
Print set Setup printing
Figure 2-89 Table 2-55
Menu Settings Comments
Fast-Cal
ON
OFF
Turn on fast-calibration
Turn off fast-calibration
Self-Cal Execute Self-calibration
Service Go to service menu
Preferenc Go to preference menu
Note:
Fast-Cal: When turn on the fast-cal function, oscilloscope will calibrate vertical
offset several times according to the running time. The Fast-Cal will run immediately
when it’s turned on.
Self-Cal: Oscilloscope will calibrate parameter of vertical system (CH1, CH2, and
Ext), horizontal system and trigger system.