
RIGOL
© Copyright RIGOL Technologies, Inc. 2007.
User’s Guide for DS1000 Series
6-4
Trigger
Trigger Sensitivity 0.1div-1.0div (adjustable)
Internal ±12 divisions from center of screen
EXT ±1.6V
Trigger Level Range
EXT/5 ±8V
Internal ±(0.3div×V/div)(±4 divisions from center of
screen)
EXT ±(6% of setting + 40 mV)
Trigger Level Accuracy
(typical) applicable for
the signal of rising and
falling time ≥20ns
EXT/5 ±(6% of setting + 200 mV)
Normal mode: pre-trigger (262144/ sampling rate),
delayed trigger 1s
Trigger Offset
Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Trigger Holdoff range 100ns-1.5s
Set Level to 50%
(Typical)
Input signal frequency ≥50Hz
Edge Trigger
Edge trigger slope Rising, Falling, Rising + Falling
Pulse Trigger
Trigger condition
(>, <, =) Positive pulse, (>, <, =) negative pulse
Pulse Width range 20ns – 10s
Video Trigger
Video standard &
line frequency
Support standard NTSC, PAL and SECAM broadcast systems.
Line number range: 1-525 (NTSC) and 1-625 (PAL/SECAM)
Slope Trigger
Trigger condition
(>, <, =) Positive slope, (>, <, =) negative slope
Time setting 20ns – 10s
Alternate Trigger
Trigger on CH1 Edge, Pulse, Video, Slope
Trigger on CH2 Edge, Pulse, Video, Slope
Pattern Trigger **
Trigger mode D0 – D15 select H, L, X, ,
Duration Trigger **
Trigger Type D0 – D15 select H, L, X
Qualifier
>, <, =
Time setup 20ns – 10s
Commentaires sur ces manuels