User’s Guide for DS1000 Series
To Set Up the Sampling System
As Figure 2-66 displays, the Acquire button for sampling system is at the MENU
Use Acquire button to pop up the menu as follows:
Table 2-41
Menu Settings Comments
Acquisition
Normal
Average
Peak Detect
Normal Acquisition mode
Average Acquisition mode
Peak Detect Acquisition mode
Averages 2 to 256
Step by multiple of two. Set
average times from 2 to 256
Sampling
Real-Time
Equ-Time
Real-time sampling mode
Equivalent sampling mode
Mem
Depth
Long Mem
Normal
Set up memory as 512K or 1M
Set up memory as 1K or 2K
Sa Rate 100.0MSa Display sampling rate
Figure 2-66
Acquire setup button
Figure 2-67