2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-281
:TRIGger:CAN:SPOint
Syntax
:TRIGger:CAN:SPOint <spo>
:TRIGger:CAN:SPOint?
Description
Set or query the sample point (expressed in percentage) of CAN trigger.
Parameter
Name Type Range Default
<spo> Integer 5 to 95 50
Return Format
The query returns the sample point of CAN trigger in integer.
:TRIGger:CAN:FTYPe
Syntax
:TRIGger:CAN:FTYPe <ftype>
:TRIGger:CAN:FTYPe?
Description
Set ro query the frame type of CAN trigger (when the trigger condition is FTYPe).
Parameter
Name Type Range Default
<ftype> Discrete {DATA|REMote|ERRor|OVERload} DATA
Explanation
When the CAN trigger condition is set to FTYPe (frame type) (
:TRIGger:CAN:WHEN), you can use this
command to set the oscilloscope to trigger when the specified type of frame is detected.
Return Format
The query returns DATA, REM, ERR or OVER.