2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-309
:TRIGger:RS232:BUSer
Syntax
:TRIGger:RS232:BUSer <user baud>
:TRIGger:RS232:BUSer?
Description
Set or query the user-defined baud rate in RS232 trigger. The unit is bps.
Parameter
Name Type Range Default
<user baud> Integer 110 to 20000000 9600
Return Format
The query returns the user-defined baud rate.
:TRIGger:RS232:DATA
Syntax
:TRIGger:RS232:DATA <data>
:TRIGger:RS232:DATA?
Description
Set or query the data value in RS232 trigger when the trigger condition is Data.
Parameter
Name Type Range Default
<data> Integer 0 to 2
n
-1 70
Note: In the expression 2
n
-1, n is the current data bits (refer to the :TRIGger:RS232:WIDTh command).
Explanation
To set the trigger condition of RS232 trigger, refer to the
:TRIGger:RS232:WHEN command.
Return Format
The query returns the data value in RS232 trigger when the trigger condition is Data in integer.