Rigol DS2000A Series Spécifications Page 305

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 390
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 304
2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-289
:TRIGger:DURAtion:WHEN
Syntax
:TRIGger:DURATion:WHEN <when>
:TRIGger:DURATion:WHEN?
Description
Set or query the trigger condition in duration trigger.
Parameter
Name Type Range Default
<when> Discrete {GREater|LESS|GLESs} GREater
Explanation
GREater: you need to specify a time. The oscilloscope triggers when the duration of the pattern is
greater than the preset time.
LESS: you need to specify a time. The oscilloscope triggers when the duration of the pattern is lower
than the preset time.
GLESs: you need to specify an upper limit of the time and lower limit of the time. The oscilloscope
triggers when the duration of the pattern is lower than the preset upper limit of the time and greater
than the preset lower limit of the time.
Return Format
The query returns GRE, LESS or GLES.
Vue de la page 304
1 2 ... 300 301 302 303 304 305 306 307 308 309 310 ... 389 390

Commentaires sur ces manuels

Pas de commentaire