2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-71
:CALibrate Commands
:CALibrate:DATE?
:CALibrate:STARt
:CALibrate:TIME?
:CALibrate:QUIT
:CALibrate:DATE?
Syntax
:CALibrate:DATE?
Description
Query the date of the last calibration.
Return Format
The query returns the date in <year>,<month>,<day> format. Wherein, <day> and <month> are
double-digit figures and <year> is a four-digit figure. For example, 2013,10,27.
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
The oscilloscope starts to execute self-calibration.
Explanation
The self-calibration can make the oscilloscope quickly reach its optimum working state to obtain the
most accurate measurement values.
The functions of most of the keys are disabled during the self-calibration.