9 Pass/Fail Test RIGOL
DS2000 User’s Guide
9-3
the two cursor lines respectively. Or, press CursorAB and use to adjust the
positions of cursor A and curcor B at the same time. Note that you can press down
continuously to switch the current cursor.
Press X Mask and Y Mask, rotate and the mask lines appear on the screen as
shown in the figure below. Press Create Mask to apply the mask currently created.
The horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and 0.03 div to
4.0 div respectively.
Test and Ouput
Before the test, you can use the following method to set the output mode of the test
results.
Press Msg Display to selectON” or “OFF. When ONis selected, the test results
will be displayed at the upper right corner of the screen.
Press Stop On Outp to select ON” or “OFF.
ON: when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOPstate. At this point, the results of the test remain the same on