Rigol DS2000 Series Guide de l'utilisateur Page 64

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 264
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 63
RIGOL 3 To Set the Horizontal System
DS2000 User’s Guide
3-8
Roll Mode
In this mode, the waveform scrolls from the right to the left to update the display and
the waveform horizontal position and trigger control are not available. The range of
horizontal scale adjustment is from 200.0 ms to 1.000 ks.
Note: when Roll mode is enabled, Delayed Sweep, Protocol Decoding”,
Pass/Fail Test, “Measurement Range, “Waveform Record, “To Set the
Persistence TimeandTo Trigger the Oscilloscopeare not available.
Slow Sweep
Another mode similar to Roll mode. When the horizontal time base is set to 200
ms/div or slower, the instrument enters slow sweepmode in which the instrument
first acquires the data at the left of the trigger point and then waits for trigger
event. After the trigger occurs, the instrument continues to finish the waveform at
the right of the trigger point. When slow sweep mode is used to observe low
frequency signal, it is recommended that the Channel Coupling is set to “DC.
Vue de la page 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 263 264

Commentaires sur ces manuels

Pas de commentaire