Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-99
:TRIGger:PATTern:LEVel
Syntax :TRIGger:PATTern:LEVel <chan>,<level>
:TRIGger:PATTern:LEVel? <chan>
Description Set or query the trigger level of the specified channel in pattern trigger. The unit is the
same as the current amplitude unit.
Parameter
Name
Type
Range
Default
<chan>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
<level> Real
(± 5 × VerticalScale from the screen center)
OFFSet
0
Explanation For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:PATTern:LEVel CHANnel2,0.16 /*Set the trigger level of CH2 to 160mV*/
Related
Command
:CHANnel<n>:SCALe
:CHANnel<n>:OFFSet