:TRIGger:RUNT:WLOWer
Syntax :TRIGger:RUNT:WLOWer <NR3>
:TRIGger:RUNT:WLOWer?
Description
Set or query the pulse width lower limit in runt trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3> Discrete 8ns to 9.99s 1μs
Explanation This command is only available when the qualifier (refer to :TRIGger:RUNT:WHEN) is
GREater or GLESs.
Return
Format
The query returns the pulse width lower limit in scientific notation.
Example
:TRIGger:RUNT:WLOWer 0.01 /*Set the pulse width lower limit to 1ms*/
:TRIGger:RUNT:WLOWer? /*The query returns 1.000000e-03*/
Related
Command
:TRIGger:RUNT:WUPPer
:TRIGger:RUNT:WHEN
:TRIGger:RUNT:ALEVel
Syntax :TRIGger:RUNT:ALEVel <level>
:TRIGger:RUNT:ALEVel?
Description Set or query the trigger level upper limit in runt trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
Type
Range
Default
<level> Real
(± 5 × VerticalScale from the screen center) OFFSet
2V
Explanation For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the trigger level upper limit in scientific notation.
Example
:TRIGger:RUNT:ALEVel 0.16 /*Set the trigger level upper limit to 160mV*/
:TRIGger:RUNT:ALEVel? /*The query returns 1.600000e-01*/