Rigol DS1000Z Series Spécifications Page 39

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 172
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 38
Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-25
:CURSor:XY
:CURSor:XY commands can only be used when the horizontal timebase mode is XY.
Command List:
:CURSor:XY:AX
:CURSor:XY:BX
:CURSor:XY:AY
:CURSor:XY:BY
:CURSor:XY:AXValue?
:CURSor:XY:AYValue?
:CURSor:XY:BXValue?
:CURSor:XY:BYValue?
:CURSor:XY:AX
Syntax :CURSor:XY:AX <x>
:CURSor:XY:AX?
Description
Set or query the horizontal position of cursor A in XY cursor measurement mode.
Parameter
Name
Type
Range
Default
<x>
Integer
5 to 394
100
Explanation In XY timebase mode, the horizontal and vertical positions are defined by the pixel
coordinate of the XY display area. The pixel coordinate of the XY display area ranges from
(0,0) to (400,400). Wherein, (0,0) is located at the right top corner and (400,400) is
located at the left bottom corner. The horizontal and vertical pixel ranges are both from 0
to 400.
Return
Format
The query returns an integer between 5 and 394.
Example
:CURSor:XY:AX 200
:CURSor:XY:BX
Syntax :CURSor:XY:BX <x>
:CURSor:XY:BX?
Description
Set or query the horizontal position of cursor B in XY cursor measurement mode.
Parameter
Name
Type
Range
Default
<x>
Integer
5 to 394
300
Explanation In XY timebase mode, the horizontal and vertical positions are defined by the pixel
coordinate of the XY display area. The pixel coordinate of the XY display area ranges from
(0,0) to (400,400). Wherein, (0,0) is located at the right top corner and (400,400) is
located at the left bottom corner. The horizontal and vertical pixel ranges are both from 0
to 400.
Return
Format
The query returns an integer between 5 and 394.
Example
:CURSor:XY:BX 200
Vue de la page 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 171 172