RIGOL Chapter 2 Command System
2-66 DS1000Z Programming Guide
[:SOURce[<n>]]:OUTPut[<n>][:STATe]
Syntax [:SOURce[<n>]]:OUTPut[<n>][:STATe] <bool>
[:SOURce[<n>]]:OUTPut[<n>][:STATe]?
Description Turn on or off the output of the specified source channel, or query the status of the
specified source channel.
Parameter
[<n>] Discrete {1|2}
If omitted, operating on
Explanation [:SOURce[<n>]] and [<n>] are used to specify the source channel. The priority of the
former is higher than that of the latter, namely if [:SOURce[<n>]] is omitted, the channel
is specified by [<n>] and if [:SOURce[<n>]] is not omitted, the channel is specified by
[:SOURce[<n>]]. If both are omitted, the commands operation on source 1 by default.
Return
The query returns OFF or ON.
[:SOURce[<n>]]:OUTPut[<n>]:IMPedance
Syntax [:SOURce[<n>]]:OUTPut[<n>]:IMPedance <impedance>
[:SOURce[<n>]]:OUTPut[<n>]:IMPedance?
Set or query the impedance of the specified source channel
Parameter
[<n>] Discrete {1|2}
If omitted, operating on
Explanation OMEG: high impedance; FIFTy: 50Ω
[:SOURce[<n>]] and [<n>] are used to specify the source channel. The priority of
the former is higher than that of the latter, namely if [:SOURce[<n>]] is omitted, the
channel is specified by [<n>] and if [:SOURce[<n>]] is not omitted, the channel is
specified by [:SOURce[<n>]]. If both are omitted, the commands operation on
Return
The query returns OMEG or FIFT.