Rigol DS1000Z Series Spécifications Page 70

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 172
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 69
RIGOL Chapter 2 Command System
2-56 DS1000Z Programming Guide
:MEASure:SETup:MAX
Syntax :MEASure:SETup:MAX <value>
:MEASure:SETup:MAX?
Description Set or query the upper limit of the threshold when measuring time, delay and phase
parameters. The default unit is %.
Parameter
Name
Type
Range
Default
<value>
Integer
7 to 95
90
Explanation The upper limit, middle point and lower limit of the threshold are used to define the
time, delay and phase parameters. Setting these values will affect the measurement
results of the time, delay and phase parameters.
Setting the upper limit lower than the middle point will automatically reduce the
middle point to keep it lower than the upper limit.
Return
Format
The query returns an integer between 7 and 95.
Example
:MEASure:SETup:MAX 95
Related
Command
:MEASure:SETup:MID
:MEASure:SETup:MIN
:MEASure:ITEM
:MEASure:SETup:MID
Syntax :MEASure:SETup:MID <value>
:MEASure:SETup:MID?
Description Set or query the middle point of the threshold when measuring time, delay and phase
parameters. The default unit is %.
Parameter
Name
Type
Range
Default
<value>
Integer
6 to 94
50
Explanation The upper limit, middle point and lower limit of the threshold are used to define the
time, delay and phase parameters. Setting these values will affect the measurement
results of the time, delay and phase parameters.
The middle point must be lower than the upper limit and larger than the lower limit.
Return
Format
The query returns an integer between 6 and 94.
Example
:MEASure:SETup:MID 89
Related
Command
:MEASure:SETup:MAX
:MEASure:SETup:MIN
:MEASure:ITEM
Vue de la page 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 171 172

Commentaires sur ces manuels

Pas de commentaire