Rigol DS1000Z Series Spécifications Page 58

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 172
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 57
RIGOL Chapter 2 Command System
2-44 DS1000Z Programming Guide
:MATH:OPTion:THReshold1
Syntax :MATH:OPTion:THReshold1 <thre>
:MATH:OPTion:THReshold1?
Description
Set or query the threshold level of the source A in logic operation. The default unit is V.
Parameter
Name
Type
Range
Default
<thre> Real
(-4*Vertical Scale Vertical Offset)~
(4*Vertical Scale Vertical Offset)
0.00V
Explanation This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Vertical Scale refers to the vertical scale of source A
Vetical Offset refers to the vertical offset of source A
The step is Vertical Scale/25
Return
Format
The query returns the threshold level in scientific notation.
Example
:MATH:OPTion:THReshold1 0.8 /* Set the threshold level to 800mV*/
:MATH:OPTion:THReshold1? /*The query returns 8.000000e-01*/
Related
Command
:CHANnel<n>:SCALe
:CHANnel<n>:OFFSet
:MATH:OPTion:THReshold2
Syntax :MATH:OPTion:THReshold2 <thre>
:MATH:OPTion:THReshold2?
Description
Set or query the threshold level of the source B in logic operation. The default unit is V.
Parameter
Name
Type
Range
Default
<thre>
Real
(-4*Vertical Scale Vertical Offset)~
(4*Vertical Scale Vertical Offset)
0.00V
Explanation This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Vertical Scale refers to the vertical scale of source B
Vetical Offset refers to the vertical offset of source B
The step is Vertical Scale/25
Return
Format
The query returns the threshold level in scientific notation.
Example
:MATH:OPTion:THReshold2 0.5
Related
Command
:CHANnel<n>:SCALe
:CHANnel<n>:OFFSet
Vue de la page 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 171 172

Commentaires sur ces manuels

Pas de commentaire