Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-81
:TIMebase[:MAIN]:SCALe
Syntax :TIMebase[:MAIN]:SCALe <scale_value>
:TIMebase[:MAIN]:SCALe?
Description
Set or query the main timebase scale. The default unit is s/div.
Parameter
Name
Type
Range
Default
<scale_value> Real Normal: 5ns to 50s in 1-2-5 step
Roll: 200ms to 50s in 1-2-5 step
s
Explanation When the horizontal timebase is YT and the timebase scale is 20ms or larger (namely
Slow Sweepmode), this command is invalid when the oscilloscope in Stoping.
Return Format
The query returns the main timebase scale in scientific notation.
Example :TIMebase:MAIN:SCALe 0.0002 /*Set the main timebase scale to 200μs*/
:TIMebase:MAIN:SCALe? /*The query returns 2.0000000e-04*/
Related
Command
:TIMebase:MODE
:TIMebase:MODE
Syntax :TIMebase:MODE <mode>
:TIMebase:MODE?
Description
Set or query the mode of the horizontal timebase.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{MAIN|XY|ROLL}
MAIN
Return Format
The query returns MAIN, XY or ROLL.
Example
:TIMebase:MODE XY
Related
Command
:TIMebase:DELay:OFFSet
:TIMebase:DELay:SCALe
:TIMebase[:MAIN]:OFFSet
:TIMebase[:MAIN]:SCALe